Principle:
The ASO-Turbilatex is a quantitative turbidimetric test for the detection of anti-steptolysin “O” (ASO) antibodies in human serum/plasma. Latex particles coated with steptolysin “O” are agglutinated when mixed with samples containing ASO. The agglutination causes an absorbance change, dependent upon the ASO contents of the patient sample that can be quantified by comparison from a calibrator of known ASO concentration.
Specification
| Methodology | Immunoturbidimetry |
| Reaction Type | Kinetic |
| Reaction Direction | Increasing |
| Sample Type | Human Serum or Plasma |
| R1:R2 Mixing Ratio | 4:1 |
| Primary Wavelength | 540 nm |
| Secondary Wavelength | NA |
| Blank Settings | Distilled Water |
| Blank Absorbance Limit | <0.300 |
| Reagent Volume Required Per Test | 1000 µl |
| Sample Volume Required Per Test | 10 µL |
| Delay Time | 10 seconds |
| Calibrator Available in the Kit | Yes |
| Factor | NA |
| Linearity | 800 IU/ml. |
| Automation | Enabled |
| Pack Sizes | 1×50 ml |
| Kit components | Reagent bottles – R1 and R2. 1 Calibrator vial. Instructions for Use. |
| Storage Temperature | 2- 80C |
| Shelf life | 24 months |
| Kit dimensions | 1×50 ml – 85x105x45 mm |
| Approx. weight of kit | 1×50 ml – 130 gms |
| Production Capacity | 600 mn ml per annum |